Posted on

Sterlink Routine Maintenance Guide

If you need to print this PDF or save for later
click below to download your copy

[button size=”medium” style=”secondary” text=”Download PDF” link=”https://vetovationnew.instawp.xyz/wp-content/uploads/2021/01/Microsoft-PowerPoint-Maintenance-Slides-1.pdf” target=”none”]

 

Basic Maintenance STERLINK

Sterlink Gas Plasma Sterilizer (FPS-15sp)

12 Month Maintenance

Sterlink Maintenance Guide | VetOvation

 

Maintenance Kits

Filters & Oil Change

Filter Kit:

Sterlink Filter Kit | VetOvation

A. HEPA Filter

B. Chamber Filter

C. Ozon Filter (Two hand knobs)

D. VC Cartridge

 

Oil Kit:

Sterlink Oil Kit | VetOvation

A. New oil injection silicon adapter (white)

B. New oil injection syringe

C. New oil bottle

D. New oil injection tube (transparent)

E. Extracted oil injection silicon adapter (black)

F. Extracted oil syringe

G. Extracted oil bottle

H. Waste oil extraction tube (black)

 

Oil Replacement Sterlink Sterilization Kit

Sterilization Kit Oil Refills | VetOvation

A. New oil injection silicon adapter (white)

B. New oil injection syringe

C. New oil bottle

D. New oil injection tube (transparent)

E. Extracted oil injection silicon adapter (black)

F. Extracted oil syringe

G. Extracted oil bottle

H. Waste oil extraction tube (black)

Oil Indicator Strip Guide | VetOvation

 

Oil Change Using Oil Replacement Kit

Sterilization Kit Oil Contents | VetOvation

This is the oil changing kit containing 2 syringe, 2 tube, 2 tube connector, 1 oil, 1 empty bottle.

Plasma Sterilizer Tube Connectors | VetOvation

Plasma Sterilizer Syringes | VetOvation

Assemble the rubber with the tube. After that, connect it to the syringe.

Sterlink Maintenance Guide | VetOvation

With your black tube, connect it to the bottom blue hold from the unit.

Sterlink System Maintenance Steps | VetOvation

After the tube is fully connected, slowly pull out the oil from the unit.

Sterlink Oil Extraction for Maintenance | VetOvation

With the empty bottle, remove the tap.

Oil Extraction for Sterlink System Maintenance | VetOvation

Pick up the black tube syringe and push to put the old oil inside the bottle.

Thorough Oil Extraction for Sterlink Maintenance | VetOvation

Repeat this until all the oil from the bottle is empty.

Sterlink Oil Refill Process | VetOvation

With a white tube syringe, take out the new oil from the bottle.

Sterlink Oil Gauge Information | VetOvation

The oil gauge should be slightly higher from the center, just like the picture.

 

Filter Replacement Sterlink Kit

Sterlink Kit Filter Replacement Contents | VetOvation

A. HEPA Filter

B. Chamber Filter

C. Ozon Filter (Two hand knobs)

D. VC Cartridge

Filter Quality Indicator Guide | VetOvation

 

Filter Change Using Filter Replacement Kit

Changing HEPA Filter

HEPA Filter Installation for Sterlink Sterilizer | VetOvation

1. Take out the new HEPA filter

Plasma Sterilizer HEPA Filter Removal | VetOvation

2. Remove the old HEPA filter by turning counterclockwise.

Plasma Sterilizer HEPA Filter Replacement | VetOvation

3. Replace it with new HEPA filter by turning clockwise

Changing Ozone Filter

Ozone Filter Location | VetOvation

1. Ozone filter can be found on the left side of the product.

Ozone Filter Replacement - Knob Removal | VetOvation

2. Remove the knobs

Ozone Filter Removal and Replacement | VetOvation

3. Remove the Ozone filter and replace it with new Ozone filter

 

STERLINK Oil Replacement Kit

[table id=3 /]

STERLINK Filter Replacement Kit

[table id=4 /]