If you need to print this PDF or save for later
click below to download your copy
[button size=”medium” style=”secondary” text=”Download PDF” link=”https://vetovationnew.instawp.xyz/wp-content/uploads/2021/01/Microsoft-PowerPoint-Maintenance-Slides-1.pdf” target=”none”]
Basic Maintenance STERLINK
Sterlink Gas Plasma Sterilizer (FPS-15sp)
12 Month Maintenance
- Change your filters and oil every 12 months, or when indicated by color.
- Use Sterlink Maintenance Kits to change Sterlink Kit filters and oils for Sterlink Sterilization Kits āAvailable for purchase from VetOvation.
- If you have any issues, send an email to VetOvation: Garrett@vetovation.com
Maintenance Kits
Filters & Oil Change
Filter Kit:
A. HEPA Filter
B. Chamber Filter
C. Ozon Filter (Two hand knobs)
D. VC Cartridge
Oil Kit:
A. New oil injection silicon adapter (white)
B. New oil injection syringe
C. New oil bottle
D. New oil injection tube (transparent)
E. Extracted oil injection silicon adapter (black)
F. Extracted oil syringe
G. Extracted oil bottle
H. Waste oil extraction tube (black)
Oil Replacement Sterlink Sterilization Kit
A. New oil injection silicon adapter (white)
B. New oil injection syringe
C. New oil bottle
D. New oil injection tube (transparent)
E. Extracted oil injection silicon adapter (black)
F. Extracted oil syringe
G. Extracted oil bottle
H. Waste oil extraction tube (black)
Oil Change Using Oil Replacement Kit
This is the oil changing kit containing 2 syringe, 2 tube, 2 tube connector, 1 oil, 1 empty bottle.
Assemble the rubber with the tube. After that, connect it to the syringe.
With your black tube, connect it to the bottom blue hold from the unit.
After the tube is fully connected, slowly pull out the oil from the unit.
With the empty bottle, remove the tap.
Pick up the black tube syringe and push to put the old oil inside the bottle.
Repeat this until all the oil from the bottle is empty.
With a white tube syringe, take out the new oil from the bottle.
The oil gauge should be slightly higher from the center, just like the picture.
Filter Replacement Sterlink Kit
A. HEPA Filter
B. Chamber Filter
C. Ozon Filter (Two hand knobs)
D. VC Cartridge
Filter Change Using Filter Replacement Kit
Changing HEPA Filter
1. Take out the new HEPA filter
2. Remove the old HEPA filter by turning counterclockwise.
3. Replace it with new HEPA filter by turning clockwise
Changing Ozone Filter
1. Ozone filter can be found on the left side of the product.
2. Remove the knobs
3. Remove the Ozone filter and replace it with new Ozone filter
STERLINK Oil Replacement Kit
[table id=3 /]
STERLINK Filter Replacement Kit
[table id=4 /]